2.5
Emergencies and errors
2.5.1
Emergency Stop button
Caution
Only push the Emergency Stop button when there is NO other choice.
Whenever possible, use EXTRACTION to release the probe from the tissue
and use the standard shutdown procedure.
Excessive use of the Emergency Stop button may damage the system.
Warning
If a major electrical risk is detected during the cryotherapy procedure, PUSH
the Emergency Stop button immediately.
2.5.2
Emergencies causing procedure halt
Warning
When a procedure halts due to an error, switch OFF the system. Call IceCure
Medical and describe the error shown on the screen as precisely as possible.
Do NOT attempt to reuse the system before contacting IceCure Medical.
Warning
When an error message is displayed and a procedure is aborted, remove the
probe ONLY after you are allowed to do so by the system. Failure to do so
will increase the risk of a liquid nitrogen related accident.
Warning
In case of software crash, switch OFF the mechanical ON/OFF button and
unplug the electrical cable.
Ca
ll IceCure Medical for technical service before restarting the ProSense™
cryoablation system.
Warning
When the system shuts itself down due to an error, contact IceCure Medical
and describe the error message shown on the screen as precisely as possible.
Do not attempt to reuse the system before contacting IceCure Medical. After