5.6
System failures
5.6.1
ProSense™
cryoablation system failure
When the
ProSense™
cryoablation system detects an
error
, the following will occur:
•
A
failure message
will appear.
•
The procedure will be aborted.
When the above occurs, take action as follows:
1.
Write down the error message and number, follow the system instructions.
Figure 112: System Fail example screen
2.
If you are in the middle of a cryotherapy procedure,
wait for passive THAW
and
then carefully
remove the cryoprobe
from the tissue.
Replace the probe when
required automatically by the system.
3.
Detach the cryoprobe
from the cryohandle
ONLY
after the system instructs you
that it is safe to do so.
4.
If system shut down is required,
shut OFF the
ProSense™
cryoablation system
by
pressing the mechanical ON/OFF button ON the main chassis and then remove
the plug from the power supply.