Figure 104: Replacement was completed after the THAW ended
5.3.8.1
Replace the Dewar during Freez e
In case of low Nitrogen,
the system can’t continue to Freeze.
It will ask to refill the
Dewar in Freeze cycle (see fig. below)
Figure 105: Replacement during Freeze, in case of empty Dewar
.
•
If you press OK, you can replace a Dewar and continue the procedure. The system will
move to Manual mode and wait in a THAW cycle for instruction.
Pay attention that in Manual mode, you need to press the icon on the screen for each
freeze or THAW cycle.
•
Replace Dewar as in paragraph 5.3.8.
•
If you press CANCEL, you will get a message of "Low Nitrogen" (See a figure below).