![IceCure ProSense FAS3000000 Скачать руководство пользователя страница 93](http://html1.mh-extra.com/html/icecure/prosense-fas3000000/prosense-fas3000000_user-manual_610296093.webp)
IceCure™ Medical Ltd.
DMS-7064 rev. C
ProSense™
Confidential
User Manual
93
European User Manual
•
Within a given protocol, if you want to increase/decrease the preprogrammed
time of a step when the step has already started, use the +/- icons on the left side
of the screen to add/subtract 15 seconds with each push.
•
At the end of the last
Freeze
step in automatic mode, the
Extraction
step will
begin and the
Extraction
screen will be displayed.
•
You can move from the automatic mode AT ANY TIME by pressing the Manual
mode button on the screen, without interrupting the procedure. When in
Manual, you can’t go back to Automatic mode
.
o
If you choose to do so, proceed as explained in section
“
Manual freeze
mode
”
: press
Next
on the touch screen or the Action button on the handle
(optional) to move from
Freeze
to
Thaw
and monitor the process under
ultrasound or other imaging modality.
Figure 87: Automatic Mode screen: press on Start to begin the Freeze step.
Figure 88:
Freeze screen in Automatic mode
During the preprogrammed procedure, the system provides information regarding
treatment progress: