IceCure™ Medical Ltd.
DMS-7064 rev. C
ProSense™
Confidential
User Manual
107
European User Manual
Figure 112: System Shutdown screen.
Press OK to proceed.
At the end of the last treatment of the day, Switch OFF the mechanical ON/OFF switch
and unplug the electrical cable. Clean the system following instructions in section 8.1 and
move the system to its storage location, using the dedicated handle.
General advice for patients
Instruct patients as follows:
•
Remove gauze dressing after 24 hours and expect minimal discharge.
•
Adhesive skin closure should be removed after 7 days if it has not already fallen off
•
They may need over-the-counter pain relievers for mild discomfort after procedure
•
After breast procedures, swelling and moderate ecchymosis may be present for
several weeks
System failures
ProSense™
cryoablation system failure
When the
ProSense™
cryoablation system detects an
error
, the following will occur:
•
A
failure
message will appear.
•
The procedure will be aborted.
When the above occurs, take action as follows:
1.
Write down the error message and number, follow the system instructions.