IceCure™ Medical Ltd.
DMS-7064 rev. C
ProSense™
Confidential
User Manual
104
European User Manual
Figure 107: Procedure stopped due to Low Liquid Nitrogen.
If you press
OK
, you will get an "Extraction Completed screen" with 3 options: Extraction, Relocate
and Finish.
Figure 108: Extraction Completed screen.
Post-operational stages
Removing the temperature sensor from tissue
At the end of the cryotherapy procedure the temperature sensor should be removed
from the tissue by carefully pulling on it. The extraction should be done only after a
passive
Thaw
of a few minutes, since the temperature sensor can be surrounded by ice.
Removing it too quickly before Thawing can damage the target tissue.
Warning
Before removing the temperature sensor from the tissue, make sure the
freeze effect has been deactivated and the temperature sensor can be easily
withdrawn. Never use excessive force to extract the temperature sensor.