IceCure™ Medical Ltd.
DMS-7063 rev. C
ProSense™
Confidential
User Manual
46
US User Manual
Warning
Cryoprobes, temperature sensors, introducers, and sterile sleeves are single-
use devices supplied in single use sterile packaging.
Reprocessing single-use devices affects their performance and can cause
cross-contamination.
Warning
After finishing the cryoablation procedure on a patient, remove and safely
discard the single-use cryoprobe, temperature sensor, introducers, and
sterile sleeve.
3.3.5.1
Handle plug holder and handle plug
The handle plug holder is located on the top of the main chassis and allows a place for
keeping the plug while performing the procedure.
Humidity is a crucial issue to address in cryosurgery systems.
After every procedure, close the handle plug on the Cryohandle.
Perform functional test at close proximity to the procedure treatment.
Figure 31: Handle plug holder and handle plug.
Figure 32: Cryohandle closed with the plug on the Handle holder
When using a 90 degrees cryohandle, pay attention that the cryohandle is directed
towards the system when laid down on the handle plug holder on main chassis.
Holder
–
Optional (Not available in some regions, e.g. China)
The Holder is an optional accessory (Not available in some regions, e.g. China), that
stabilizes the hose in a specific position during the procedure.
The holder is adjustable in height and length and should be adjusted by the user as
required.