IceCure Medical Ltd.
DSR3000000 rev.L
Confidential
User Manual
66
IceSense3™
5.3.4
Pause Option:
Pause option is available at any freeze cycle, in Manual or in Automatic mode. The use of the button
is according to clinical judgment ( for re-positioning the probe in the tissue or injecting saline, for
example). Using the pause option should be as minimal as possible. While using the pause button,
freezing mode is disabled therefore is not calculated in the total freezing cycle time.
Figure 55: Pause and Play buttons during the freeze cycle.
The pause time does not affect the freeze preprogrammed time but depending on when and how
long the pause time is defined it may affect the iceball size.
The physician should always monitor the iceball size during the whole procedure.
IceCure Medical recommends is to use pause only if necessary.
Pause
Play