User Manual
59
IceSense3™
Figure 47: Functional test - visual inspection screen
If a functional problem occurs or there is any unusual appearance (such as frost on the plastic
cover near the cryohandle, foam or any unusual appearance) and it is not represented by one
of the Illustrations, press "Cancel". Follow system instructions until you are required to safely
remove the probe from the cryohandle. You will then be returned to the Main Menu screen.
To start again Check that the Dewar is full and replace the probe. If the problem recurs, do
not proceed with treatment. Turn off the system and contact IceCure Medical.
If a functional problem occurs as represented by one of the pictures, press on the relevant
picture.
Follow system instructions until you are required to safely remove the probe from the
cryohandle.
You will then be returned to the Main Menu screen
.
Check that the Dewar is full and replace
the probe and start again. If the problem recurs do not proceed with treatment. Turn off the
system and contact IceCure Medical.
Choosing ”Success” will load the treatment selection screen.
If for any other reason you decide to stop the procedure, press “Cancel”. Follow system
instructions until you are required to safely remove the probe from the cryohandle. You will
then be returned to the Main Menu screen.
Warning
In case of frost on shaft, start active Warm if possible. If not, wait for passive Thaw. In both cases
use skin protection techniques.