IceCure Medical Ltd.
DSR3000000 rev.L
Confidential
User Manual
58
IceSense3™
Figure 46: Functional test screen
The system will check essential parameters.
A failure of any internal test will result in, an error message displayed on screen. Please record
the error number and follow system instructions until you are requested to remove the probe
safely from the cryohandle. You will then be returned to the Main Menu screen. Contact
IceCure Medical technical service.
After successful internal test, the system will display the visual inspection screen. You will be
prompted to inspect the following:
o
The cryoprobe tip – make sure a small ice ball forms.
o
The cryoprobe shaft – check that there is no ice on it.
o
The saline – check that there are no bubbles.
Illustrations of potential functional failures will appear on screen. Please refer to these
illustrations while inspecting the cryoprobe.