User Manual
56
IceSense3™
5.2.7.4
Cryoprobe connection
Connect the cryoprobe to the cryohandle as follows, while maintaining sterility of the probe:
Remove the plug that covers the probe connection point.
Insert the cryoprobe into the insertion point in the handle as shown on screen and screw it
until a “Next” button appears on screen to confirm that probe is in place.
Remove the probe cover that protects the tip.
When done, press the Action button on the handle or “Next” on the touch screen.
Figure 45: Cryoprobe connection screen
Warning
Never reuse a single-use cryoprobe or a single-use sterile sleeve.
Warning
Never “unscrew” the cryoprobe if you are not clearly allowed to unscrew or disengage it.
After pressing Next, a “Do not disconnect” message will appear
Warning
After completion of treatment of a given patient, make sure that the single-use cryoprobe
and single-use sterile sleeve are removed and discarded.