Chapter 10
– Malfunctions
209
Process was canceled
manually
Cleaning and disinfection process was
stopped by user
Cleaning and disinfection process
was not completed!
Program must be restarted
301
Overflow switch active
Unit is leaking (e.g. water, wash,
processing media)
Turn off water supply!
Be sure to inform specialist
supplier/customer service!
Fluids in the unit, e.g. in the base pan,
may contain processing media in case
of damage. Remember to follow the
safety precautions of the processing
media manufacturer
302
Fresh water supplied even
though inlet valve is closed
Inlet valve is defective
Immediately turn off water supply!
303
Overheating
Electronics are defective
Contact customer service
304
Temperature too low
Heating element defective
Contact customer service
305
Circulation pump defective Foreign bodies in pump wheel
306
Temperature sensor -
Heating element
No water
Temperature sensor defective
Check water supply
Contact customer service
307
Temperature sensor -
Chamber
Temperature sensor defective
Contact customer service
308
Internal error
Software error
Missing component
Turn off unit and restart
Call customer service