Chapter 10
– Malfunctions
General situation
Possible causes
Recommended course of action
Remaining moisture on
instruments
Base rack and/or basket inserts
or racks not loaded properly or too
full
Door was closed for a long time after
program ended (see logs)
Arrange washware properly and do
not overload
Dry instruments manually
Open the door immediately after program ends.
Remaining moisture in
hollow instruments
Dry narrow-necked instruments manually
with sterile compressed air