26
Chapter 5
– Cleaning and disinfecting
Start time preselection
The thermal disinfector HD450 and HD450 Injection has a freely programmable time
delay.
To optimize your processing time, you can select the free start time between 1 and 99
hours in the hour rhythm.
The thermodesinfector then starts automatically at the desired time.
For a delayed program start, select "Start time".
Now you can select any start time.
After selecting the start time, confirm the entry with "OK" and enter a 3-digit
code. Once the start time has elapsed, the thermodisinfector starts