OPERATION MANUAL 

Revision: January 2021 

 
 
 
 
 

 
 
 

ECHO-FLEX

™ 5002 

 

 

 

 
 

Содержание ECHO-FLEX 5002

Страница 1: ...N MANUAL ECHO FLEX 5002 IBIOM Instruments Ltd 670 Joseph Louis Mathieu Street Sherbrooke Quebec J1R 0X3 Tel 1 877 563 5777 Fax 450 445 9837 info ibiom com OPERATION MANUAL Revision January 2021 ECHO F...

Страница 2: ...12 6 4 Table functions 13 6 5 Security rail and cardiac cutout 15 6 6 Positioning cushion 17 6 7 IV pole 18 6 8 Rechargeable battery 19 6 9 Cardiac cutout in backrest 20 7 CLEANING INSTRUCTIONS 21 7 1...

Страница 3: ...have the following information ready 7 digit serial number example 1234567 Model number or name example 5002 This information is available on label located on base of the table Phone 1 877 563 5777 F...

Страница 4: ...a situation which if not avoided could result in death or serious injury It may also describe potential serious adverse reactions and safety hazards Warning electrical shock hazard Operator must read...

Страница 5: ...205kg on table Safe working load includes patient s weight and accessories WARNING Never exceed safe working load of 150lb or 68kg on cardiac cutout WARNING Never exceed safe working load of 150lb or...

Страница 6: ...y rail is closed make sure it is securely locked WARNING Once cardiac cutout is closed make sure it is securely locked WARNING Once positioning cushion is in place make sure it is securely locked WARN...

Страница 7: ...s name and address Product code or name 7 digit serial number Product s weight Date of manufacture YYYY MM DD Maximum safe working load maximum patient weight Made in Canada QR code pointing to IBIOM...

Страница 8: ...ill beeps hold down 2 buttons together of Trendelenburg adjustment 3rd row at the same time until it stops beeping It will reset all errors in control box b Control box will beep but table will not mo...

Страница 9: ...ith defibrillator WARNING Do not roll on power cable hand or foot control cable WARNING Keep all cables out of reach of children WARNING Keep all cables away from moving parts WARNING Do not move tabl...

Страница 10: ...OPERATION MANUAL ECHO FLEX 5002 Copyright IBIOM Instruments ltd 2021 Page 10 of 31 Revision January 2021...

Страница 11: ...2 Once in place install all accessories IV pole arm board if applicable 3 Once in place lock all wheels 4 Remove any remaining packaging from the product Example Protective plastic on the hand control...

Страница 12: ...ficient clearance around the table before activating the various movements to avoid collisions with the environment equipment and people WARNING Do not use ECHO FLEX table as a transport stretcher to...

Страница 13: ...NING Medical staff should inform the patient on how to adjust table WARNING If patient is not able to adjust table the hand control should be placed out of the patient s reach WARNING Avoid placing ha...

Страница 14: ...e foot control Button Function A Move up table B Move down table C Tilt up back rest section D Tilt down back rest section Triple foot control Button Function A Move up table B Move down table C Tilt...

Страница 15: ...re it is securely locked WARNING Once cardiac cutout is closed make sure it is securely locked WARNING Always raise security rail once the patient is in place WARNING Do not leave patient unattended W...

Страница 16: ...PERATION MANUAL ECHO FLEX 5002 Copyright IBIOM Instruments ltd 2021 Page 16 of 31 Revision January 2021 Security rail lowered down Cardiac cutout opened Cardiac cutout opened Security rail lowered dow...

Страница 17: ...m WARNING Once positioning cushion is in place make sure it is securely locked WARNING Do not leave patient unattended To lift positioning cushion pull handle Move up cushion to one of the 3 positions...

Страница 18: ...7 IV pole WARNING Always ensure there is sufficient clearance around the table before activating the various movements to avoid collisions with the environment equipment and people WARNING Never exce...

Страница 19: ...ry may develop excessive heat CAUTION NEVER discharge completely battery or it may be damaged and reduce its lifetime Always allow battery to settle to room temperature before use Do not exceed duty c...

Страница 20: ...body parts near moving parts or locking mechanism WARNING Once cardiac cutout is closed make sure it is securely locked WARNING Do not leave patient unattended WARNING Always have patient enter and ex...

Страница 21: ...ther products or in concentration other than those specified in this section WARNING Do not use abrasive cleaning products WARNING Avoid placing hands and other body parts near moving parts or locking...

Страница 22: ...l mattress surfaces including surfaces in openings between the backrest and main section IMPORTANT Allow the surfaces to dry before raising the backrest o NOTE In case of major spill it is possible to...

Страница 23: ...ckrest 0 to 80 Adjustable height 21 to 37 in 53 to 94 cm Trendelenburg 0 to 15 Reverse Trendelenburg 0 to 15 Security rail length 24 in 61 cm Maximum safe working load 450 lb 205 kg Wheel diameter 5 i...

Страница 24: ...tery capacity option 1 3 Ah Protection class IP Control box IPX6 Actuator IPX4 Lifting column IPX6 Hand control IPX6 Double foot switch IPX6 Triple foot switch IP20 Button controller IPX0 Battery IPX6...

Страница 25: ...IBIOM Instruments ltd 2021 Page 25 of 31 Revision January 2021 8 4 Certification All IBIOM s product are certified according to CSA standard for electrical medical device except contrary indication La...

Страница 26: ...ion or by misuse of the user guide instructions and warnings Also it doesn t apply if modifications or repairs have been done by someone not authorized by IBIOM instruments ltd Electrical components a...

Страница 27: ...dispose of it properly in accordance with the regulations in force in your region Before disposal equipment must be clean properly to avoid contamination Equipment can be disposed of by separating th...

Страница 28: ...OPERATION MANUAL ECHO FLEX 5002 Copyright IBIOM Instruments ltd 2021 Page 28 of 31 Revision January 2021 11 MAINTENANCE See service manual...

Страница 29: ...the movements If problem persists go to next cause CAUTION Never allow the battery to discharge completely as this may damage the battery Green light on the control box with or without audible alarm...

Страница 30: ...OPERATION MANUAL ECHO FLEX 5002 Copyright IBIOM Instruments ltd 2021 Page 30 of 31 Revision January 2021 13 SPARE PARTS LIST See service manual...

Страница 31: ...RATION MANUAL ECHO FLEX 5002 Copyright IBIOM Instruments ltd 2021 Page 31 of 31 Revision January 2021 IBIOM INSTRUMENTS LTD 670 JOSEPH LOUIS MATHIEU STREET SHERBROOKE QU BEC J1R 0X3 CANADA www ibiom c...