Gas Incubation System
• Only ibidi technical staff and technical staff instructed by ibidi are permitted to open and ser-
vice the Gas Incubation System.
• The external power supply should not be brought into contact with moisture. If the housing is
damaged, the external power supply should not be used.
• Avoid strong magnetic fields and sources of high frequency. The Gas Incubation System might
not function properly when located near a strong magnetic field or high frequency source.
• Avoid vibrations from vacuum pumps, centrifuges, electric motors, processing equipment, and
machine tools.
• Avoid dust and corrosive gas. Do not install the Gas Incubation System where it could be
exposed to high levels of dust or to outside air or ventilation outlets.
• Install the Gas Incubation System in a horizontal and stable position, such as a table, bench, or
desk upon which the instrument is installed.
• Install the Gas Incubation System in a location that enables easy access for maintenance.
• Do not place heavy objects on the instrument.
• The Gas Mixer needs a gas input of 1 bar/14.5 psi. Only operate the Gas Mixer with the pressure
indicated in the specifications on page
1.7
Regulatory Statement
The Gas Incubation System has been designed, produced and tested in compliance with the European
standard DIN EN 61010-1 (IEC 61010-1, ”Safety requirements for electrical equipment for measure-
ment, control and laboratory use”). Furthermore it meets the IEC 61326-1 (”Electrical equipment
for measurement, control and laboratory use - EMC requirements”) and CISPR 11 (”International
Standard for electromagnetic emissions (disturbances) from Industrial, Scientific and Medical (ISM)
Equipment”) standards .
The device carries the CE mark.
The Gas Incubation System meets the Low Voltage Directive 2014/35/EU and the
EMC Directive 2014/30/EC.
Instruction Manual
11