IBI
SCIENTIFIC
9861 Kapp Court Peosta, IA 52068
Tele:
(563) 690-0484
Toll Free:
1-800-253-4942
Fax:
(563) 690-0490
E-Mail:
Web:
www.ibisci.com
IBI SCIENTIFIC
ROLOM0002 – Rev. 1
ISO 9001:2008 Certified
ROLMT0001
– Replacement Roller Motor Installation
Important!
– Prior to starting project, ensure unit is unplugged from electrical source.
1.)
First, using a phillips screw driver loosen the (2) flat-head screws located on the front lower part
of the front cover. See balloon “A”. Once those (2) screws have been removed, locate the pan-
head screws on each side of the unit. There is (1) screw on each side, located near the front of the
unit. Remove these (2) screws as well. See balloon “B”.
2.)
Now carefully rotate the front cover forward and place on its face in front of the unit. This will
expose the inside of the unit.
NOTE:
Cables connecting the PC boards and motor assembly are
short in length, use caution not to damage cables or connectors when disassembling.
3.)
Now remove two wire (red/black) wire harness from the PC board, located on the left side of the
base of the unit. This will help free the front cover from the base of the unit. See balloon “D”.
Next locate the ribbon cable from the motor to the PC board mounted to the underside of front
cover. Using a small flat screw driver, carefully slide the two tabs outward on the connector
mounted on the PC Board (where the motor cable is connected). Once the two tabs have been
slid outward the motor cable will freely slide out of the connector. See balloon “E”. This will
free the front cover from the base of the unit. Carefully place front cover out of the way for now.
See balloon “C”.