![IBA PADU-8-ICP Скачать руководство пользователя страница 40](http://html1.mh-extra.com/html/iba/padu-8-icp/padu-8-icp_manual_3237246040.webp)
Manual ibaPADU-8-ICP
40 Issue
1.7
11
Technical data
Order number iba
10.120100
Mechanical stability and
test parameters (all 3 axes)
DIN IEC 68-2-6;
1 g rms 90 Min @ 0..250 Hz / axis (all types)
2 g rms 90 Min @ 0..250 Hz / axis (all ibaPADU-8 types);
ibaPADU-32-R 2 g`s with additional rear holder only !
EMI test parameters
EN 55011 (Class A); EN61000-4-6 (Class 3); EN61000-4-3/ENV
50204 (Class 3)
Operating temperature
0
°
C to 50
°
C (32 °F…122 °F),
Storage temperature
-25
°
C to 70
°
C (-13 °F…158 °F)
transport temperature
-25
°
C to 70
°
C (-13 °F…158 °F)
Cooling
Self cooling
Mounting
On DIN-rail, snap-on
Humidity Class
F no moisture
Protection Class
IP20
Power supply
24 V DC +/-20 % non stabilized
Power/Current consumption (w/o load)
typ. 300 mA, max. 400 mA; I
OFF->ON
approx. 1A
Sampling rate programmable up to
25 kHz 8 channels simultaneously
FO-cable
Coupling
62,5/125 µm
ST Lean
Max. length of fiber optics between devices
2000 m (6560 ft)
Dimensions (WxHxD) [mm] ([inch])
54 x 194 x 155 (2.13 x 7.64 x 6.10)
Weight (incl. packaging and documentation)
1050 g (37 oz)
Analog inputs
Number
8
Each channel has its own ADC
Resolution
14 Bit
Input level / type
ICP Sensor input +/-5 V, with
3mA current loop
Max. length of sensor cable 30 m
Dig. Low pass filter
6-pole, 72 dB
Low-pass corner frequency from 0.5
Hz to 12.5 kHz, set by software
High pass input characteristic
0.5 Hz / 6 dB
0.1 Hz / 6 dB on request
Due to the ICP measuring principle
DC measurement is impossible
Programmable input gain
0..60 dB
In steps of 0.5 dB
Dynamic range SNR
84 - 140 dB
Galvanic Isolation
Channel / Dev.-ground
1.5 kV
Binary inputs
Number
8
Input level
log 0; < +/-6 V
log 1; > +/-8 V
Sampling
Simultaneously sampled with
analog inputs
max. 25 kHz /channel
Galvanic isolation
Channel/Channel/Digital
Ground
1.5 kV