IACER Srl
47
MNPG118-02
PROBLEM
POSSIBLE CAUSE
SOLUTION
Electrodes not correctly
placed on the skin.
Check the electrodes position
according to electrodes
positions user manual. The
distance between the
electrodes should be at least
of 5 cm.
The stimulation is
ineffective.
Improper electrodes.
Reposition or replace the
electrodes.
Unknown.
Consult the doctor.
Display shows the
error E1 or E2.
Hardware problem.
Restart the device, if the
problem persists contact the
fabricant.
Display shows the
error E3.
Temperature sensor failure.
The device will stop treatment
automatically, please wait at
least 30 minutes before
restart treatment.
Display shows the
error E4.
Excessive temperature of
internal parts.
The device will stop treatment
automatically, please wait at
least 30 minutes before
restart treatment.
Display shows the
error E5.
Internal memory error.
Restart the device, if the
problem persists contact the
fabricant.
The stimulation is
ineffective.
Improper electrodes.
Reposition or replace the
electrodes.
Some commands
don’t work
regularly.
Defective keys and buttons.
Contact the manufacturer
Failure electronic control
circuit.
The device works
properly, but there
is a noticeable
drop in the
effectiveness of
the treatment.
Possible head failure.
Contact the manufacturer
Possible failure of the
device current generator
circuit.
Disposal
The I-TECH UE, device was designed and engineered to have minimal negative
environmental impact, in consideration of its performance and safety