handpiece and the symbol on the display will start flashing.
22.
It’s possible to stop temporary the therapy at any time pressing the
knob
(3). Press again the knob to continue the treatment
23.
Press the
orange button
to stop immediately the treatment in
progress.
ATTENTION
.
For safety patient, the device is equipped with a
protection system against high temperatures. The device will
stop electrical stimulation when the feature board temperature
reaches 80° C. The device cannot work again unless the
temperature is below 60°C.
Moreover, for patient safety, the device will stop ultrasound
treatment and LED starts flashing on the handle if the
temperature is above 42°C. It will start again when the
temperature goes down to 41°C.
NOTE: for more information’s about electrodes refer to the previous paragraph
Electrodes
.
Reset
To reset parameters and return the device to the factory settings on the device
follow the instructions below:
•
switch off the device;
•
keeping pressed simultaneously the
knobs
of electrotherapy channels
1 and 2 (1/2);
•
switch on the device by using
ON/OFF
button;
•
keeping pressed the
knobs
(1) and (2): the device emits a prolonged
signal for around 5 seconds and the set-up frame will showed on the
display.
In this way the device is reset and all the custom settings are eliminated.