Use in the electrotherapy mode
Operative instructions
Follow the indications below to start the combined treatment.
1.
In combo mode ultrasound probe works as negative of channel 2, you
should connect electrotherapy cable with 1 derivation (red cable) to the
green cable on the output 2.
The channel 1 is turned off.
2.
Connect the green cable to the
socket
(12) placed on the back side of the
device.
3.
Connect the
electrode
to the
red cable
for electrotherapy.
4.
Place the electrode on the treatment area following the instruction of the
previous paragraphs. The electrode should be placed near the painful
zone, approximately 10/15 cm far from the painful point where the
ultrasound head is moved.
5.
Connect the ultrasound handle to the
socket
(11) placed on the back side
of the device. Connect the cable when the device is switched off.
6.
Switch on the device pressing
ON/OFF
button placed laterally, next to
power supply socket. IT IS RECOMMENDED TO TURN OFF THE DEVICE
BEFORE CONNECTING ALL THE CABLES!
7.
Immediately after switching on, the device carries out a self-test (around
10 seconds). At the end of the self-test display shows the last performed
treatment.
8.
Press
WAVEFORM MODE (B1)
button to show
: this icon indicates
the COMBINED ULTRASOUND/ELECTROTHERAPY treatment mode.
9.
Select waveforms by using the
knob
(3): IF-2P, TENS, EMS and RUSSIAN
(Kotz).
10.
As indicated in the paragraph related to treatment with electrotherapy,
each waveform has 10 stored programs (see the related tables on the
previous pages). Select the program by pressing the
B2 PROGRAM CC / CV
button and turn the
knob
(3) on the chosen program. As in the
electrotherapy mode, also in the combined mode it is possible to choose
between single-phase P or three-phase S programs, by keeping pressed
PROGRAM CC / CV
button for at least 5 seconds.