HyQuest Solutions HyQual Скачать руководство пользователя страница 17

 

HyQual Multi-probes User Manual v3, 08.07.2022 
KISTERS Controlled, Template: Word-with-table-of-contents-Template-Company.docx, v1.1 

Page 17 / 23 

 

7.3.7

 

Conductivity 

HyQuest Solutions uses the four-electrode method to determine water conductivity.  Two pairs of 
graphite electrodes are situated in a stable geometry (you can barely see the electrodes; they look like 

two bull’s eyes inside the slot on the conductivity sensor).

 

A constant voltage is applied to one of each electrode pair, and the amount of current required to main-
tain that voltage is measured.  As the conductivity of the water increases, the current increases predicta-
bly. 
Conductivity sensor maintenance is nothing more than occasionally cleaning the measurement surface 
with a soft cloth or cotton swab and soapy water.  Do not use anything abrasive. 

The zero point for the sensor is set electronically, so you need only set the “slope” point:

 

1.

 

Open the Calibration menu and select 

“SpCond µS/cm” or “SpCond mS/cm”, depending on your 

preference of units. 

2.

 

Screw your calibration cup onto the HyQual probe housing and remove the cup lid.  Rinse your 

sensors several times with the standard you’ll use for calibration.

 

3.

 

With the sensors pointed upward, fill the calibration cup until your solution covers the sensors 
by an inch or so. 

4.

 

Wait a few minutes for the temperature to equilibrate and the sensor to arrive at a steady read-
ing. 

5.

 

Now just follow the calibration instructions on the screen. 

 
The HyQual probe normally reports Specific Conductance 

 

that’s Conductivity that has been standard-

ized to 25°C.  Your reading is thus the conductivity of your water if that water was exactly 25°C.  Conduc-
tivity has three other forms, Total Dissolved Solid

s (TDS), Salinity, and Resistivity (which we don’t report 

because it has little meaning in natural waters).  You can’t calibrate TDS or salinity directly because they 
are calculated from Conductivity.  You can, however, “calibrate” TDS with a TDS standard 

by adjusting 

the conductivity calibration point up or down until the TDS standard produces the desired TDS reading.  

The same is true for Salinity if you’re using a standard quantified on the Practical Salinity Scale (PSS).  

"Enable" TDS and/or Salinity by checking the box next to those parameters in the "Sensors and Parame-
ters" section. 

7.3.8

 

pH 

pH is measured as the voltage drop across the glass membrane of a pH electrode.  A reference electrode 
is used to complete the voltage-measuring circuit.  The pH glass is specially formulated to absorb water 
so that ions (particularly H+ and OH-) in the water are attracted to the glass to offset the ionic constitu-

ency of the pH electrode’s internal electrolyte.  As a result, there is a charge separation across the glass, 
and that’s the voltage we measure.  pH readings are automatically compensated for temperature.

 

pH electrode maintenance is nothing more than occasionally cleaning the glass surface with a soft cloth 
and soapy water.  Do not use anything abrasive.  The important part of pH maintenance is refilling the 
reference electrode.  (D.10) 
You can choose a two- or three-point pH calibration.  The two-point calibration, a seven buffer and a 
second buffer whose value is near that of the waters you intend to monitor, is recommended.  If you are 
measuring in waters whose pH might range significantly above and below seven, you may be able to in-
crease your accuracy slightly by choosing a three-point calibration (seven buffers plus one basic buffer 
and one acidic buffer).  pH calibration is simple: 

1.

 

Open the Calibration menu and select “pH”.

 

2.

 

Screw your calibration cup onto the HyQual probe housing and remove the cup lid.  Rinse your 

sensors several times with the standard you’ll

 use for calibration. 

Содержание HyQual

Страница 1: ...HyQual Water Quality Multi Probe HyQuest Solutions KISTERS Group 2022 05 03...

Страница 2: ...ction 10 6 2 2 Unattended Logging 10 6 2 3 Telemetry Relay 11 6 2 4 On Line Monitoring 12 7 1 When to proceed to maintenance 12 7 2 Maintenance 1 Keep your device clean 12 7 3 Maintenance 1 Calibratio...

Страница 3: ...ilidad de tales da os Cualquier error encontrado en cualquier producto de HyQuest Solutions debe ser comuni cado a HyQuest Solutions donde se har todo lo posible para resolver r pidamente el pro blema...

Страница 4: ...industrial specification that defines the electrical interface and physical layer for point to point communication of electrical devices The RS 485 stand ard allows for long cabling distances in elec...

Страница 5: ...s battery operates only at the surface and its job is to power the multiprobe while you re pairing the instrument with a data display via Bluetooth wireless technology A USB port on a PC is another wa...

Страница 6: ...send the probe back to us at your own expense Additional sensors Barometric pressure is calculated with the use of the depth sensor therefore you will need to add an optional depth sensor to any of yo...

Страница 7: ...ufacturer s named repair center You can find information on how to re turn items for repair in the relevant section of the HyQuest Solutions website Disposal instructions After taking any HyQuest Solu...

Страница 8: ...ufficient water flow that can be easily ac cessed for maintenance and protected from external elements such as animals and items such as stones or sticks that may damage the probe 4 1 Protection for u...

Страница 9: ...n to toggle between Logging ON and Logging OFF The Setup function for Logging lets you name your Logging file and set the logging interval time be tween logged lines of data Figure 1 Switch Logging ON...

Страница 10: ...The Display can be a laptop or almost any tablet or smart phone You can record measurements using the HyQual probe s Snapshot or Automatic Snapshot features You might snapshot a series of measurements...

Страница 11: ...rchase this battery after as it is not integrated to the probe Please note that this battery ads several inches to the length of the probe and that it can also be used with the HyQual 200 only that it...

Страница 12: ...ed in the control loop or decision making process Maintenance 7 1 When to proceed to maintenance Judgment gained from observing your field conditions and data requirements provides information about w...

Страница 13: ...tuation for which the correct parameter value is known Here s the general procedure 1 Clean the sensor and perform any necessary sensor specific maintenance 2 Select a calibration standard whose value...

Страница 14: ...c But you can elect to accept any SRF 7 3 3 Choosing Calibration Standards For best results choose a calibration standard whose value is close to what you expect to see in the field For example calibr...

Страница 15: ...ate of the calibration the parameter calibrated the reading before the calibration was accepted the read ing after the calibration was accepted the raw sensor reading the SRF and a few other details I...

Страница 16: ...follow the calibration instructions on the screen The manufacturers of optical DO sensors usually recommend that you not calibrate the zero DO point because the zero DO point drift is very low However...

Страница 17: ...don t report because it has little meaning in natural waters You can t calibrate TDS or salinity directly because they are calculated from Conductivity You can however calibrate TDS with a TDS standar...

Страница 18: ...ous circle at the end of the sleeve This not only purges bubbles from the electrolyte but also cleans nasty stuff out of the junction Figure 4 Filling in fresh pH reference electrolyte 7 3 10 ORP ORP...

Страница 19: ...s a small hole in the back of the transducer There is a tube inside the cable i e a Vented Cable to connect the Stage sensor to the atmosphere so that changes in barometric pressure will not affect t...

Страница 20: ...e small set screw on the wiper arm 2 Remove the wiper pad from the wiper arm and replace the pad 3 Place a new wiper arm on the motor shaft so that the set screw faces the flat spot on the motor shaft...

Страница 21: ...bor charge 1 2 days down time Repair HyQuest Solutions precision instruments and data loggers are produced in quality controlled processes All HyQuest Solutions production and assembly sites in Austra...

Страница 22: ...lay devices not for eld use calibration solutions copper anti fouling kit USB adapter connect DB9 to USB inte grated SDI 12 and MODBUS output SDI 12 adapter cable MODBUS adapter cable and Android disp...

Страница 23: ...ion battery must be dis posed of according to legal provisions EU WEEE Directive As players in the environmental market KISTERS AG and HyQuest Solutions are committed to supporting efforts to avoid a...

Отзывы: