HyQuest Solutions HyQual Скачать руководство пользователя страница 14

 

HyQual Multi-probes User Manual v3, 08.07.2022 
KISTERS Controlled, Template: Word-with-table-of-contents-Template-Company.docx, v1.1 

Page 14 / 23 

 

5.

 

Access your app’s calibration function by navigating from the Home Screen to 

the calibration 

section.  Select the parameter to be calibrated, and then enter the calibration value in the type-
in box, and press enter.  When the reading has stabilized, press enter to calibrate. The HyQual 
probe will report the resulting Sensor Response Factor (SRF; see below). For most apps, you 
then press Y to accept the calibration, N to back up one step, or Exit to leave the sensor uncali-
brated. 

7.3.2

 

Sensor Response Factor (SRF) 

Near the end of the calibration routine, you will be asked to 
accept or decline the calibration based on the Sensor Re-

sponse Factor (SRF).  Suppose that a “typical” Conductivity 
sensor puts out 100 μA in a 1413 μS/cm standard. If your 
Conductivity sensor reports 100 μA in that same calibration solution, then your SRF is 100% (some

 pa-

rameters, such as pH, have a more complex SRF calculation, but the effect is the same).  If your response 

is 80 μA, your SRF would be 80%.  When you press the OK button to accept a calibration, 

the HyQual 

probe automatically accepts your calibration if the SRF is between 60% and 140%.  If the SRF falls out-
side that range, you will be cautioned to check your standard value, make sure the sensor is clean, make 
sure the reading has stabilized, etc. But you can elect to accept any SRF. 

7.3.3

 

Choosing Calibration Standards 

For best results, choose a calibration standard whose value is close to what you expect to see in the 

field. For example, calibrate with a 1413 μS/cm Specific Conductance standard if you expect to see Spe-

cific Conductance readings between 500 and 

1000 μS/cm in the field. Calibrating with a sea water stand-

ard or a very low standard would not be appropriate. Similarly, if your waters tend toward the acidic, 
calibrate with a 4-buffer instead of a 10-buffer. 
If you are moving The HyQual probe across wide ranges of water conditions, you may wish to recalibrate 
to match the new situations.  For instance, if you are measuring a clear lake during the morning and a 
high-sediment stream in the afternoon, you might consider recalibrating at noon with a high-value tur-
bidity standard. 
The table below shows common calibration practices. 

Table 1 

 Common Calibration Practices 

Sensor 

Standard Method of Calibration 

Available Calibration Solutions 

Comments 

Temperature 

never requires calibrating 

N/A 

 

pH / pH reference 

2 or 3 points 

pH 4, pH 7, pH 10 

pH7, pH 10 most common 

ORP 

1 point 

ORP Standard 200 mV 

 

Conductivity 

1 point 

CD Standard, 0.5 Molar, 58670 Micro S 
CD Standard, 0.1 Molar, 12856 Micro S 
CD Standard, 0.01 Molar, 1412 Micro S 
CD Standard, 0.001 Molar,147 Micro S 

brackish/saltwater 

borderline brackish typical 

freshwater very pure 

fresh/glacial 

Reference Electrode 

calibration not required 

N/A 

replace pH electrolyte 

solution at routine 

calibration 

Depth 

adjust for barometric pressure 

N/A 

recalibrate at deployment 

site for best accuracy 

 

Содержание HyQual

Страница 1: ...HyQual Water Quality Multi Probe HyQuest Solutions KISTERS Group 2022 05 03...

Страница 2: ...ction 10 6 2 2 Unattended Logging 10 6 2 3 Telemetry Relay 11 6 2 4 On Line Monitoring 12 7 1 When to proceed to maintenance 12 7 2 Maintenance 1 Keep your device clean 12 7 3 Maintenance 1 Calibratio...

Страница 3: ...ilidad de tales da os Cualquier error encontrado en cualquier producto de HyQuest Solutions debe ser comuni cado a HyQuest Solutions donde se har todo lo posible para resolver r pidamente el pro blema...

Страница 4: ...industrial specification that defines the electrical interface and physical layer for point to point communication of electrical devices The RS 485 stand ard allows for long cabling distances in elec...

Страница 5: ...s battery operates only at the surface and its job is to power the multiprobe while you re pairing the instrument with a data display via Bluetooth wireless technology A USB port on a PC is another wa...

Страница 6: ...send the probe back to us at your own expense Additional sensors Barometric pressure is calculated with the use of the depth sensor therefore you will need to add an optional depth sensor to any of yo...

Страница 7: ...ufacturer s named repair center You can find information on how to re turn items for repair in the relevant section of the HyQuest Solutions website Disposal instructions After taking any HyQuest Solu...

Страница 8: ...ufficient water flow that can be easily ac cessed for maintenance and protected from external elements such as animals and items such as stones or sticks that may damage the probe 4 1 Protection for u...

Страница 9: ...n to toggle between Logging ON and Logging OFF The Setup function for Logging lets you name your Logging file and set the logging interval time be tween logged lines of data Figure 1 Switch Logging ON...

Страница 10: ...The Display can be a laptop or almost any tablet or smart phone You can record measurements using the HyQual probe s Snapshot or Automatic Snapshot features You might snapshot a series of measurements...

Страница 11: ...rchase this battery after as it is not integrated to the probe Please note that this battery ads several inches to the length of the probe and that it can also be used with the HyQual 200 only that it...

Страница 12: ...ed in the control loop or decision making process Maintenance 7 1 When to proceed to maintenance Judgment gained from observing your field conditions and data requirements provides information about w...

Страница 13: ...tuation for which the correct parameter value is known Here s the general procedure 1 Clean the sensor and perform any necessary sensor specific maintenance 2 Select a calibration standard whose value...

Страница 14: ...c But you can elect to accept any SRF 7 3 3 Choosing Calibration Standards For best results choose a calibration standard whose value is close to what you expect to see in the field For example calibr...

Страница 15: ...ate of the calibration the parameter calibrated the reading before the calibration was accepted the read ing after the calibration was accepted the raw sensor reading the SRF and a few other details I...

Страница 16: ...follow the calibration instructions on the screen The manufacturers of optical DO sensors usually recommend that you not calibrate the zero DO point because the zero DO point drift is very low However...

Страница 17: ...don t report because it has little meaning in natural waters You can t calibrate TDS or salinity directly because they are calculated from Conductivity You can however calibrate TDS with a TDS standar...

Страница 18: ...ous circle at the end of the sleeve This not only purges bubbles from the electrolyte but also cleans nasty stuff out of the junction Figure 4 Filling in fresh pH reference electrolyte 7 3 10 ORP ORP...

Страница 19: ...s a small hole in the back of the transducer There is a tube inside the cable i e a Vented Cable to connect the Stage sensor to the atmosphere so that changes in barometric pressure will not affect t...

Страница 20: ...e small set screw on the wiper arm 2 Remove the wiper pad from the wiper arm and replace the pad 3 Place a new wiper arm on the motor shaft so that the set screw faces the flat spot on the motor shaft...

Страница 21: ...bor charge 1 2 days down time Repair HyQuest Solutions precision instruments and data loggers are produced in quality controlled processes All HyQuest Solutions production and assembly sites in Austra...

Страница 22: ...lay devices not for eld use calibration solutions copper anti fouling kit USB adapter connect DB9 to USB inte grated SDI 12 and MODBUS output SDI 12 adapter cable MODBUS adapter cable and Android disp...

Страница 23: ...ion battery must be dis posed of according to legal provisions EU WEEE Directive As players in the environmental market KISTERS AG and HyQuest Solutions are committed to supporting efforts to avoid a...

Отзывы: