Hypersen Technologies Co., Ltd.
HPS-FT-USER V1.8
28
/
38
Precautions:
1. By default, the sensor enables crosstalk compensation mode, which has a strong inhibitory
effect on crosstalk, but will slightly affect the single-axis precision.
2. If the operating scenario is single-axis measurement, you can try to disable the crosstalk
compensation mode.
5.1.8 Overload information receiving function
Steps:
1. Connect the host software to the sensor according to 5.1.1.
2. Click the
Stop
button to stop the data acquisition.
3.
Click
Setting
to enter the overload information receiving interface.
4. You can get the sensor overload times (counted once every 20ms based on the measuring range of
the sensor, where the count is updated once every 2min) and overload peak (which may be
abnormal if the sensor is damaged).
5.1.9 Voltage check function
Steps:
1. Connect the host software to the sensor according to 5.1.1.
2.
Click the
Stop
button to stop the acquisition performed by the sensor.
3.
Click
Setting
to enter the
Display
interface.
4. Click the
Raw Voltage
to check the voltage value (the calculation method will change and it is
recommended to clear it after switching). Click
Force
to hide the voltage value.
5.
Return to the “Plot” interface, and click
Start
to start acquiring data.