![Hydronix Hydro-Probe XT Скачать руководство пользователя страница 31](http://html.mh-extra.com/html/hydronix/hydro-probe-xt/hydro-probe-xt_user-manual_2192332031.webp)
Electrical Installation and Communication
Hydro-Probe XT User Guide HD0538 Rev: 1.6.0 31
6 Connecting to a PC
A converter is required to connect the RS485 interface to a PC. Up to 16 sensors may be
connected at any time.
RS485 line termination will not normally be required in applications with up to 100 m of cable.
For longer lengths connect a resistor (approximately 100 Ohm) in series with a 1000pF capacitor
across each end of the cable.
It is highly recommended that the RS485 signals be run to the control panel even if they are
unlikely to be used as this will facilitate the use of diagnostic software should the need arise.
There are four types of converter supplied by Hydronix.
6.1
RS232 to RS485 converter – D type (part no: 0049B)
Manufactured by KK systems, this RS232 to RS485 converter is suitable for connecting up to
six sensors on a network. The converter has a terminal block for connecting the twisted pair
RS485 A and B wires and can then be connected directly in to the PC serial communication
port.
1
2
3
4
5
RS232
RS485
K2
-
ADE
Connect
to PC
serial port
Hydronix Part No 0049B
6 DIP Switches control the configuration of the converter.
For both the 0049 and 0049B these should be set to:
Switch 1 ON
Switch 2 OFF
Switch 3 OFF
Switch 4 ON
Switch 5 OFF
Switch 6 OFF
RS485 A
RS485 B
Connect to
Sensor
Cable
Figure 26: RS232/485 converter connections (0049B)
6.2
RS232 to RS485 converter – DIN rail mounting (part no: 0049A)
Manufactured by KK systems, this powered RS232 to RS485 converter is suitable for
connecting up to 16 sensors on a network. The converter has a terminal block for connecting
the twisted pair RS485 A and B wires and can then be connected to a PC serial
communication port.
Power Supply
7-35v DC
KD485-STD
1
2
3
4
5
6
1
2
3
4
5
6
P1
P2
7
8
+
-
TX
RX
RTS
GND
PC
or
Control System
RS232
(full duplex)
Host System
3
2
7
5
9-way D-Type PC
connector
1 2 3 4
Hydronix
Sensor
F RS485(A)
G RS485(B)
H
Hydronix
Sensor
F RS485(A)
G RS485(B)
H
DIP Switch
Settings
RJ11
RS232
Port
Hydronix Part No. 0049A
Figure 27: RS232/485 converter connections (0049A)
Содержание Hydro-Probe XT
Страница 4: ...4 Hydro Probe XT User Guide HD0538 Rev 1 6 0 ...
Страница 6: ...6 Hydro Probe XT User Guide HD0538 Rev 1 6 0 ...
Страница 8: ...8 Hydro Probe XT User Guide HD0538 Rev 1 6 0 ...
Страница 10: ...10 Hydro Probe XT User Guide HD0538 Rev 1 6 0 ...
Страница 14: ...Chapter 1 Introduction 14 Hydro Probe XT User Guide HD0538 Rev 1 6 0 ...
Страница 42: ...Chapter 4 Configuration 42 Hydro Probe XT User Guide HD0538 Rev 1 6 0 ...
Страница 52: ...Chapter 6 Optimising Sensor and Process Performance 52 Hydro Probe XT User Guide HD0538 Rev 1 6 0 ...
Страница 58: ...Chapter 7 Sensor diagnostics 58 Hydro Probe XT User Guide HD0538 Rev 1 6 0 ...
Страница 60: ...Chapter 8 Technical specification 60 Hydro Probe XT User Guide HD0538 Rev 1 6 0 ...
Страница 66: ...Appendix B Document Cross Reference 66 Hydro Probe XT User Guide HD0538 Rev 1 6 0 ...