30 Hydro-View User Guide HD0531 Rev 2.0.0
5 Configuring the Overview Screen
If a sensor is connected to the Hydro-View at start-up, the Overview Screen configuration
manager will be displayed. This enables the user to configure which values are displayed on the
Overview screen. If the overview screen has been previously configured and the same sensors
are connected the configuration manager will not be displayed.
The Overview screen (Chapter 5 Section 2) can be manually configured to show 1, 2 or 4 display
areas – see page 40. Each of these display areas is configurable. It is possible to select which
sensor, which reading and whether this is shown numerically or graphically. A graphical display
can be configured to display different time periods and different vertical ranges. See page 40.
Any combination is possible, so with four display areas selected it is possible to show four
readings from one sensor, one reading from four sensors, different readings from all four
sensors, or the same reading from the same sensor both numerically and graphically. The
selection of this setup is entirely down to site requirements and user preferences.
6 Calibrating the Sensors
If the Hydro-View is to show true moisture (rather than the Unscaled reading) or the control
system requires an input from the sensor that corresponds to a moisture % then the sensors
must be calibrated to the material being measured. As each sensor may be installed in a different
manner, it is advised that each is individually calibrated even if they are measuring in similar
materials. For this reason, the Hydro-View does not allow calibrations to be copied from one
sensor to another.
Detailed information regarding calibration and the calibration procedure is given in Chapter 6.
7 Upgrading the Hydro-View Firmware
From time to time Hydronix will issue updates to the Hydro-View Firmware. These may add new
features and functions to the product and are also designed to improve performance.
To upgrade the firmware, proceed as follows:
1. Download the latest version of the Firmware HS0097 from www.hydronix.com.
2. This will be a .zip file, which should be extracted onto the root of a USB Memory stick.
Doing this should create a folder called HydroView_IV\Upgrade which contains a number
of files.
3. Turn off power to the Hydro-View and plug the memory stick into the USB socket on the
side of the Hydro-View, or into the USB Extension socket if this has been installed.
4. Turn power on again, and the Hydro-View will detect the Upgrade files and automatically
upgrade itself to the new version.
5. When the upgrade is complete and the Overview screen is shown, the USB Memory stick
may be removed.
Removal of the USB Memory stick before the upgrade is complete may
damage your Hydro-View and prevent it from starting up.
Removal of power to the Hydro-View before the upgrade is complete may
damage your Hydro-View and prevent it from starting up.
8 Backing up the System
Once the Hydro-View is fully configured and the sensors calibrated, it is advisable to carry out a
system backup (Page 41) to keep a record of the configuration. This may be useful should the
unit fail and need to be replaced in the future. Note that only one backup may be stored on a
single USB memory stick, so a backup should be copied to a PC as soon as possible to prevent
it being over-written by another backup in the future. The name of the file should not be changed
if it is to be used to restore a system. Using a suitable folder structure on the PC will enable
backups from a number of Hydro-View units to be stored.
Содержание HV04
Страница 4: ...4 Hydro View User Guide HD0531 Rev 2 0 0 ...
Страница 6: ...6 Hydro View User Guide HD0531 Rev 2 0 0 ...
Страница 12: ...12 Hydro View User Guide HD0531 Rev 2 0 0 ...
Страница 14: ...14 Hydro View User Guide HD0531 Rev 2 0 0 ...
Страница 22: ...Mechanical Installation Chapter 2 22 Hydro View User Guide HD0531 Rev 2 0 0 ...
Страница 28: ...Electrical Installation Chapter 3 28 Hydro View User Guide HD0531 Rev 2 0 0 ...
Страница 78: ...Material Calibration Chapter 6 78 Hydro View User Guide HD0531 Rev 2 0 0 ...
Страница 80: ...Default PIN Codes Appendix A 80 Hydro View User Guide HD0531 Rev 2 0 0 ...
Страница 82: ...USB Memory Stick file format Appendix B 82 Hydro View User Guide HD0531 Rev 2 0 0 ...
Страница 84: ...Quick Start Rules Appendix C 84 Hydro View User Guide HD0531 Rev 2 0 0 ...
Страница 86: ...Frequently Asked Questions Appendix D 86 Hydro View User Guide HD0531 Rev 2 0 0 ...
Страница 92: ...Document Cross Reference Appendix G 92 Hydro View User Guide HD0531 Rev 2 0 0 ...