BIGboy™ Flow Meter

1

CALIBRATION

This meter has a permanent factory calibration for 

measuring water. If installed and used correctly, 

inaccuracies of no greater than ±5% will be obtained.

INSTALLATION

This meter may be installed in-line either horizontally 

or vertically. Installation to metal connections is not 

recommended. Install as follows:

1.  Apply teflon tape to all fittings (customer supplied). 
2.  Install meter in the arrow direction of flow.
3.  Hand-tighten fitting on both ends of the Flow Meter. 

DO NOT

 overtighten with tools, which will damage 

the plastic housing.

OPERATION

Batch and Cumulative Totals

The meter maintains two totals. The batch total  

may be reset to measure flow during a single use. The 

cumulative total provides continuous measurement and 

may not be manually reset. The batch total is labeled 

as TTL1. The cumulative total is labeled TTL2.

When the cumulative total reaches a maximum reading 

of 9999, it will automatically reset to zero.
Press the DISPLAY button briefly to switch between the 

batch and cumulative total.

Activate the Meter

Turn the meter ON by starting water flow or briefly 

pressing the DISPLAY button. The meter will display the 

batch or cumulative total from last use.
Press DISPLAY briefly to display the batch total. Hold 

the DISPLAY button down for three seconds to reset the 

batch total to zero.
The meter is programmed to turn off automatically if 

not used for about one minute.

MAINTENANCE

Proper handling and care will extend the life and 

service of the meter.

Turbine Rotor

The meter is virtually maintenance-free. However, it is 

important the rotor moves freely. Keep the meter clean 

and free of contaminants.
If the rotor does not turn freely, apply a penetrating 

lubricant on the rotor, shaft, and bearings. Remove any 

debris or deposits from the rotor using a soft brush or 

small probe. Be careful not to damage the turbine rotor 

or supports.

Battery Replacement

The meter is powered by two AAA alkaline batteries 

which may be replaced while the meter is installed. 

When batteries are removed or lose power, the batch 

and cumulative totals reset to zero but the factory 

calibration is retained.

Batteries included.