5.
To change the view options, push . Select an option, then push
OK
.
Note: The view options settings do not change what is recorded to log files. All measurements are recorded to
log files.
Option
Description
Parameters
Sets the parameters that show on the display. Push
OK
to select or not select a
parameter.
Parameter
Order
Sets the sequence the parameters show in the list view.
To move a parameter up or down the list, select the parameter, push
OK
and then push
the
UP
or
DOWN
arrow.
Graph Order
Sets the parameters that show in the graph view. Only the first two parameters in the list
show in graph view.
To move a parameter up or down the list, select the parameter, push
OK
and then push
the
UP
or
DOWN
arrow.
Grid Order
Sets the parameters that show in the grid view and the locations of the parameters in the
grid. Only the first four parameters in the list show in grid view.
To move a parameter up or down the list, select the parameter, push
OK
and then push
the
UP
or
DOWN
arrow.
Update Interval
Sets the time interval between measurements.
Stability Check
Sets the stability check to on or off. When set to on, a status icon shows at the bottom of
the display that identifies if the measurement is stable or not stable. Refer to
on page 10 for status icon descriptions. Measurements that are not stable are identified in
the log files.
Note: A measurement is identified as stable when it is within the selected stability criteria
for the parameter.
Stability
Criteria
Sets the stability check criteria used by the software to identify when measurements are
stable.
Select a parameter and push
OK
. Push to set:
Max Delta
—The maximum difference between the current measurement and the average
measurement for a measurement to be identified as stable.
Number of samples
—the
number of measurements that are used to calculate the average measurement used for
stability checks.
Note: Stability check settings are not used for calibrations.
6.
To go out of real-time monitoring, push .
Figure 8 Real-time monitoring views
Save real-time measurements
Real-time measurements are manually saved to a log file(s) on the instrument.
English
17
Содержание Surveyor HL
Страница 2: ...English 3 Deutsch 26 Italiano 52 Français 77 Español 102 Português 127 中文 153 日本語 174 Polski 198 2 ...
Страница 152: ...Descrição Nº de item Alça de mão 9118400 Cabo USB com conector mini B 1 m 3 pés 9070800 152 Português ...
Страница 158: ...图 4 校准线缆或设置线缆连接 图 5 通信模块 USB 接口 连接 用户界面及导航 按键说明 请参阅 图 6 了解按键说明 158 中文 ...
Страница 179: ...図 4 校正ケーブルまたは配備用ケーブルの接続 図 5 通信モジュールの接続 ユーザインターフェースとナビゲーション キーパッドの説明 キーの説明については図 6 を参照してください 日本語 179 ...
Страница 225: ......