Troubleshooting Chart:
* In hard water areas, scale may form inside the discharge end of the eductor, as well as in other areas of the unit that are
exposed to water. This scale may be removed by soaking the eductor in a descaling solution (deliming solution). To remove
an eductor located in the cabinet, firmly grasp vacuum breaker and unthread eductor. Replace in same manner.
Alternatively, a scaled eductor can be cleaned (or kept from scaling) by drawing the descaling solution through the unit.
Operate the unit with the suction tube in the descaling solution. Operate the unit until solution is drawn consistently, then
flush the unit by drawing clear water through it for a minute. Replace concentrate container and put suction tube into
concentrate.
1. No discharge
a. No water
a. Open water supply
b. Magnetic valve not functioning
b. Install valve parts kit
c. Excessive water pressure
c. Install regulator if water pressure exceeds
4.3 bar (flowing)
d. Eductor clogged
d. Clean* or replace
2. No concentrate draw
a. Clogged foot valve
a. Clean or replace
b. Metering tip or eductor has
b. Clean (descale)* or replace
scale build-up
c. Low water pressure
c. Minimum 1.43 (with water running)
required to operate unit properly
d. Discharge tube(s) not in place
d. Push tube firmly onto eductor discharge
or flooding ring missing from
hose barb; be sure inner discharge tube
inner discharge tube
is installed and has flooding ring.
e. Concentrate container empty
e. Replace with full container
f. Inlet hose barb not screwed
f. Tighten, but do not overtighten
into eductor tightly
g. Clogged water inlet strainer
g. Disconnect inlet water line and clean
strainer
h. Selector out of position
h. Assure selector is in position desired
3. Excess concentrate draw
a. Metering tip not in place
a. Press correct tip firmly into barb on
eductor
b. Chemical above eductor
b. Place concentrate below the eductor
4. Failure of unit to turn off
a. Water valve parts dirty or
a. Clean* or replace with valve parts kit
defective
b. Magnet doesn't fully return
b. Make sure magnet moves freely.
c. Push button stuck
c. Remove button and clean cabinet/button
to remove any dirt lodged in slide recess
5. Excess foaming in discharge a. Air leak in pick-up tube
a. Put clamp on tube or replace tube if brittle
b. Inner discharge tube not in place
b. Install inner discharge tube
Solution
Cause
Problem
10084450
Hydro Systems Europe Ltd
, Unit 3, The Sterling Centre, Eastern Road, Bracknell, Berks RG12 2PW UK
Tel: +44 (0)1344 48 88 80
Fax: +44 (0)1344 48 88 79
www.hydrosystemseurope.com