Setting the HART sensor address
HART sensors can be used with the Hydro-Logic
®
Flexi Logger 300Ex. The
possible address range for HART sensors is from 0 to 15. HART sensors are
often shipped with the address set to 0.
With the Hydro-Logic
®
Flexi Logger 300Ex in HART mode, the sensor address
must be in the range from
1 to 3
. If you are only using one HART sensor, the
address should be set to
2
.
With firmware 21.75 or newer, you can set the HART address using the logger
and a PC with a serial terminal program
– there is no need for a separate HART
modem. If required,
To set the HART address,
•
Connect the Hydro-Logic
®
Flexi Logger 300Ex to your PC with a serial
cable.
•
Connect ONLY the sensor you wish to set the address of to any sensor
connector on the logger.
•
Ensure that the logger is configured for at least one HART sensor.
Enable this via Change Sensor in Harvest if necessary.
•
Open a
terminal program
(examples are Putty or Tera Term)
•
Set the serial parameters to the correct port, 9600 baud 8 1 N:
•
Press the up/down arrow keys a few times. The logger should wake up
and display something like this on the terminal software screen:
Содержание Hydro-Logic Flexi Logger 300Ex
Страница 1: ...www hydro int com LoggerSupport Hydro Logic Flexi Logger 300Ex User Manual ...
Страница 65: ...Hydro Logic Flexi Logger 300Ex User Manual 22 August 2018 Page 65 of 74 Appendix 1 IECEx Certification ...
Страница 66: ...Hydro Logic Flexi Logger 300Ex User Manual 22 August 2018 Page 66 of 74 ...
Страница 67: ...Hydro Logic Flexi Logger 300Ex User Manual 22 August 2018 Page 67 of 74 ...
Страница 68: ...Hydro Logic Flexi Logger 300Ex User Manual 22 August 2018 Page 68 of 74 ...
Страница 69: ...Hydro Logic Flexi Logger 300Ex User Manual 22 August 2018 Page 69 of 74 ...
Страница 70: ...Hydro Logic Flexi Logger 300Ex User Manual 22 August 2018 Page 70 of 74 Appendix 2 ATEX Certificate ...
Страница 71: ...Hydro Logic Flexi Logger 300Ex User Manual 22 August 2018 Page 71 of 74 ...
Страница 72: ...Hydro Logic Flexi Logger 300Ex User Manual 22 August 2018 Page 72 of 74 ...