4-20 mA Sensor
One 4-20mA sensor can be connected, wired as shown in
The warm-up time, power supply and sensor ranging are the same as for HART
sensors.
4-20mA sensors
cannot be mixed
with HART sensor since the loop current is
shared and will not be accurate.
Connecting a 4-20mA sensor will prevent use of digital input 2 on the other
sensor connectors, since this pin is used as the sensor positive loop connection.
The sensor is powered for a user-specified warm-up period (1) before a
measurement is taken. Long warm-up periods will reduce battery life.