HydrINS 2 / HydrINS 2 Mini flow meter
User Manual
Page 57 / 79
11.2
Replacing the batteries of the HydrINS 2 / HydrINS 2 Mini flow meter
Apply the following procedure to replace the batteries in the head of the HydrINS 2 / HydrINS 2 Mini
flow meter:
1 – Open the head of the flow meter (see Para. 8.3).
2 – Disconnect the used battery and connect the new one.
3 – In Winfluid, click the
HydrINS II
or
Mini HydrINS
input channel. In the
Sensor
tab, click
Properties
, and then
Advanced.
In the
Maintenance
tab, select the boxes for Batteries 1 and 2 in the
head, and then confirm by clicking
OK
. The battery level will then be reinitialised during probe
programming.
4 – Click "Sensor" and then "program".
Figure 41: Reinitialisation of the battery levels in HydrINS 2 in Mode 1 in Winfluid
11.3
Replacing the display batteries
1 – Open the upper cover of the Display.
2 – Remove the used battery and insert the new battery.
3 – Close the cover of the Display.
4 – In Winfluid, click the
HydrINS II
or
Mini HydrINS
input channel. In the
Sensor
tab, click
Properties
, and then
Advanced.
In the
Maintenance
tab, select the boxes for Battery 2, indicate the
capacity in Ah of the new battery, and then confirm by clicking
OK
. The battery level will then be
reinitialised during probe programming.