HydrINS 2 / HydrINS 2 Mini flow meter
Installation Manual
Page 17 / 79
3.2
Models in the HydrINS 2 / HydrINS 2 Mini flow meter range
The following table is a selection guide to identify the most appropriate HydrINS 2 model according
to the pipe diameter and dimensions as shown in the diagram above:
A: Total stem length (see the above diagram)
B: Stacking length of HydrINS 2/HydrINS 2 Mini accessories (see dimensions on the diagram
above)
C: Stacking height at insertion level (theoretical value, must be checked on site)
HYDREKA
Catalogue
Number
Model
Hydrins
Dimension
A
mm
(inches)
Dimension
B
mm
(inches)
Dimension
C*
mm
(inches)
Available
insertion
length
= A-(B+C)
mm
(inches)
Usable on pipe
diameters
mm
(inches)
Profile Centre
SI_HYDA2P-30
300
689
(27.12)
234
(9.21)
200
(7.87)
255
(10)
< 250
(9.84)
< 500
(19.68)
SI_HYDA2P-50
500
869
(34.21)
234
(9.21)
200
(7.87)
400
(15.75)
< 400
(15.75)
< 800
(31.5)
SI_HYDA2P-70
700
1109
(43.66)
234
(9.21)
200
(7.87)
675
(26.57)
< 650
(25.59)
< 1300
(51.18)
SI_HYDA2P-
100
1000
1409
(55.47)
234
(9.21)
200
(7.87)
975
(38.38)
< 950
(37.4)
< 1900
(74.8)
SI_HYDE2-15
Mini
500 (19.68)
165
(6.5)
200
(7.87)
175
(6.89)
< 180
(7.08)
< 300
(11.81)
Table 7: Flow meter selection table according to the flow range to be measured
Note: This is a theoretical value. It is provided as an indication, but must be checked on site.
The model is also chosen according to the space available in the port. Allow sufficient distance above
the insertion point for installation in a port or valve chamber, known as the clearance height.