SMU1000 Factory
setting
HYDAC FILTER SYSTEMS GMBH
en(us)
Page 32 / 48
BeWa SMU1000 34x 3614430a en-us 2012-03-13.doc
2012-03-13
Factory setting
The basic settings when the SMU is delivered are:
-
The contrast is at the setting 0.
-
The background lighting is constantly activated.
-
Menu setting: LIGHTING ✔
Software Update
If the existing program was deleted (see controller main menu), the program is
available on the SMU support CD.
To do this, copy the files under "SMU1000Bootprojekt_Vxxx" onto the memory card.
Then disable the write-protection of the files on the MMC card. Select the files and
open the properties by right-clicking. Remove the checkmark next to "Write-
protected".
When the memory card is inserted, the program is automatically loaded again with a
system restart.
If the program has been deleted (see main menu "Control unit"), the program is still
saved on the memory card and can be loaded automatically by restarting the
system. You can also carry out a software update in this way.
Checking sensor settings (CS1000 / CS2000)
The connected sensors CS1000 and CS2000 must be set up or adjusted for
operation with the SMU in accordance with the following table.
You can find details in the corresponding sensor documentation.
The sensors AS1000 / AS2000 / AS8000 / HYDACLab and VLxGW do not need
further adjustment for operation on the SMU.
Sensor
Settings
CS1000
ANA.OUT = HDA.ISO
Measurement time = 60 sec.
CS2000
The PLC output must give a signal based on the ISO code.
MCS1000
Evaluation is only possible from switching output 1.
Configure the MCS accordingly.