MCS 1000 Series
Measuring menu
HYDAC FILTER SYSTEMS GMBH
en(us)
Page 54 / 76
BeWa MCS1000 3435427f V130 en-us 2012-06-15.doc
2012-06-15
CYC
E:
non-ferromagnetic
particles (nFe)
Class E
CYC
F:
non-ferromagnetic
particles (nFe)
Class F
STATUS: Status
byte
(00 at status = O.K.)
FI:
Field strength of the field coil in the MCS
TEMP C: Medium temperature in °C (indirect
measurement via sensor in the spool)
TEMP F: Medium temperature in °F (indirect
measurement via sensor in the spool)
Factory setting:
FE A
The value of 1000 µm is shown on the display as 1k.
SWT.OUT - Set switching output
Function:
You can find a detailed description on page 41.
S1 = Ferromagnetic (Fe)
FE.NFE
S2 = Non-ferromagnetic (nFe)
S1 = ferromagnetic (Fe) and non-
ferromagnetic (nFe)
Parameters:
ALL.RDY
S2 = Device Ready
Factory setting:
FE.NFE