FCU 8000 series
Starting up the FCU
HYDAC FILTER SYSTEMS GMBH
en(us)
Page 40/112
BEWA FCU8000 3109776d en-us 2011-06-10.doc
2011-06-10
Starting up the FCU
Conducting Measurements via the High-Pressure Port (INLET)
Press the "START" key and set the flow rate to approx. 50 ml/min. using the flow
control valve on the front.
If an error message appears in the display, turn the flow control valve in the required
direction until the error message disappears and measurement operation starts
automatically.
Operating the FCU
The various operating menus of the FCU are described below. The following
conventions are used:
-
Fixed display texts
are highlighted in light gray (printed version) or yellow (PDF
file).
-
Variable display texts
(user entries) are highlighted in
bold
(printed version) or
magenta
(PDF file)
-
Comments
are shown in
italics
.
Function of the Keys
Control buttons (gray)
ok
start
Confirm entries / Start measurement
stop
Cancel entries / Stop measurement
Scroll through menus (when the “
” menu symbol is displayed)
+
Scroll through Selection. Raise/lower numbers and letters (in
Input mode)