6.5.3 Filter membrane holder with a screw cap
To remove the filter membrane, proceed as follows:
ü
Wait until the analysis fluid has been fully extracted.
ü
End the extraction via the user interface.
ü
Shut the ball valve upstream of the filter membrane holder.
1. Hold the holder by the lower part with one hand.
Loosen the union nut counterclockwise with the other hand.
2. Put down the lower part together with the filter membrane in
its holder.
3. Remove the filter membrane for further evaluation and mark
it as shown in the chapter
.
1
2
3