9 | Troubleshooting/fault rectification
HYDAC FILTER SYSTEMS GMBH
66 / 84
BeWa CSI-C-11 4548075 from_V250-V253 en-us web
LED
High
Byte
Low
Byte
Description
Rectification
Red
0
21
An e-mail notifi-
cation could not be
sent.
Check the
- Mail server
address,
- User name,
- Password and
- Recipient's e-
mail address.
Red
0
22
CMX Socket error,
CSI cannot
establish a
connection to the
CMX.
Check the
network
connection and
the CMX server
settings.
Red
0
23
CMX Socket error,
CSI cannot
establish a
connection to the
CMX.
Check the
network
connection and
the CMX server
settings.
Red
0
24
CMX Parse error,
connection to the
server present,
data could not be
stored.
Check the CMX
data such as:
- User name,
- password,
- Station ID and
- Datapoint ID.
Red
0
25
CMX Encode error Restart the CSI.
Red
1
X
x = Error code of
sensor 1.
Read the details
in the documen-
tation for sensor
1.
Red
2
y
y = Error code of
sensor 2.
Read the details
in the documen-
tation for sensor
2.
Tab. 18: Troubleshooting
Signal light – Evaluating / reading the Error Code
To determine the current status of the signal light from the
Error Code, mask the particular bits:
1
2
3
4
5
6
7
8
9
1
0
A