CSI-B-1
CS 1000 - Reading out measured values
HYDAC FILTER SYSTEMS GMBH
en(us)
Page 26 / 36
BeWa CSI-B-1 3424452b 2012-04-20 en-us.doc
2012-04-20
CS 1000 - Reading out measured values
If a CS 1000 is detected at the sensor input, then the CSI-B-1 outputs the following
analog signals:
Output
OUT1
SAE A (0.0 … 14.0)
or
ISO >4 µm
(9.6 … 23.7)
OUT2
SAE B (0.0 … 14.0)
or
ISO >6 µm
(8.3 … 22.3)
OUT3
SAE C (0.0 … 14.0)
or
ISO >14 µm
(5.8 … 19.8)
OUT4
SAE D (0.0 … 14.0)
The values in the tables can be calculated as follows:
Value
Formula
Voltage U
= 2 + 8 x SAE class / 14
Current I
= 4 + 16 x SAE class / 14
SAE class
= (voltage U - 2) x 14 / 8
SAE class
= (current I - 4) x 14 / 16
ISO > 4µm
~ SAE A + 9.6
ISO > 6µm
~ SAE B + 8.3
ISO > 14µm
~ SAE C + 5.8
ISO > 21µm
~ SAE D + 3.3