ALPC 9000 series
Preparing sample bottles for analysis
Preparing sample bottles for analysis
Prepare the sample bottles for use as follows:
1. Check the sample bottles for sufficient
filling. For an analysis, at least 50 ml
fluid are required.
2. Shake the sample bottle vigorously for
~ 1 minute. Afterwards, the deposited
(sedimented) contamination is
dissolved and evenly distributed in the
fluid.
3. Let the sample bottle sit ~ 15 minutes
or place the bottle in an ultrasonic
bath.. This allows air bubbles which
resulted from the "Test fluid" to
escape.
HYDAC FILTER SYSTEMS GMBH
en
Page 35 / 68
BEWA ALPC9000 3601005 en 2011-04-07.doc
2011-04-06