45
HOCT-1F/1
6.4.
Anterior segment image operation (optional)
6.4.1.
Preparation for anterior segment operation
1.
Check the lens surface of anterior segment adapter is clean.
2.
Thread anterior segment adapter (A) to objective lens holder (B), and check there is no tilting or
misalignment of adopter. Don’t lock it too hard.
3.
Check anterior headrest rubber is clean.
4.
Hang anterior headrest rubber to headrest of instrument (C) first, then press anterior headrest rubber
to be fixed firmly.
5.
Check the mounted anterior headrest rubber does not incline.