![Huvitz HLM-1 Скачать руководство пользователя страница 19](http://html1.mh-extra.com/html/huvitz/hlm-1/hlm-1_service-manual_2190264019.webp)
18
AUTO
-
LE
NSMETE
R
2.8.
How to change the LED Assembly
The spare part for the LED assembly is the Upper Frame. So, we provide all upper
frame assembly for the LED assembly.
The procedures for changing the upper frame Assembly are as followings:
A.
Remove the Side Cover, Front Upper Cover and Upper Frame.
B.
Place the new Upper Frame and fasten it.
C.
Assemble the Front Upper Cover and the Side Cover again.
D.
You must execute all procedures for the calibration that are described in from the
section 2.2 to section 2.6.
2.9.
How to change the Pinhole Housing
The procedures for changing the Pinhole Housing are as followings:
1.
Take out the Lens Supporter.
2.
Remove the Pinhole Housing Cover which is the below cover under the Lens Supporter.
This is stuck with soft glue. Therefore we need to lift up smoothly.
Содержание HLM-1
Страница 1: ......
Страница 2: ...Auto Lensmeter HLM 1 1 Service Manual Auto Lensmeter HLM 1...
Страница 21: ...20 AUTO LENSMETER 3 Repair Standard 3 1 Removing Cover Assembly P I c t u r e...
Страница 23: ...22 AUTO LENSMETER 3 2 Frame Assemblely P I c t u r e...
Страница 25: ...24 AUTO LENSMETER 3 3 Back Bone Assembly P I c t u r e...
Страница 27: ...26 AUTO LENSMETER 3 3 1 PD Bar Sub Assembly P I c t u r e...
Страница 29: ...28 AUTO LENSMETER 3 3 2 Foot Frame Sub Assembly P I c t u r e...
Страница 31: ...30 AUTO LENSMETER 3 3 3 PEN Frame Sub Assembly P I c t u r e...
Страница 33: ...32 AUTO LENSMETER 3 3 4 PEN Slide Sub Assembly P I c t u r e...
Страница 35: ...34 AUTO LENSMETER 3 3 5 PD Pinion Sub Assembly P I c t u r e...
Страница 37: ...36 AUTO LENSMETER 3 3 6 LED Bone Sub Assembly P I c t u r e...
Страница 39: ...38 AUTO LENSMETER 3 4 LCD Assembly P I c t u r e...
Страница 41: ...40 AUTO LENSMETER 3 5 Under Cover Assembly P I c t u r e...
Страница 44: ...Auto Lensmeter HLM 1 43...
Страница 49: ...48 AUTO LENSMETER C Set the connected COM port and baud rate 115200 D Press Connect button...
Страница 51: ...50 AUTO LENSMETER G Wait until download is the end H Reboot the Lensmeter after restart system message...