background image

PRODUCT

Auto Ref/Keratomet

er

SERVICE MANUAL

DOC No.

SCS-I-7RK-005

 PREPARE   REVIEW

CONFIRM

DATE

2015.11.17

MODEL

CR/CRK-7000

REVISION DATE

REVISON

 

Adjust CCD camera

USE

2015.11.17

Fault

Faulty part

Cause

1. After replacing the CCD      

camera without 6 points

1. CCD camera position

Action

What

When

1. Adjust CCDC camera position

Содержание CR-7000

Страница 1: ...RK 005 PREPARE REVIEW CONFIRM DATE 2015 11 17 MODEL CR CRK 7000 REVISION DATE REVISON Adjust CCD camera USE 2015 11 17 Fault Faulty part Cause 1 After replacing the CCD camera without 6 points 1 CCD camera position Action What When 1 Adjust CCDC camera position ...

Страница 2: ...Step 1 Entering engineer mode Turn on Unit CRK 7000 Power After initial Press button in order of 4 3 4 3 6 6 Using button 4 and 5 select 4 Motor test setup Push SEL to enter button 6 1 2 3 4 5 6 Step 2 Choose Alignment Select 1 Alignment content and press ENTER button 6th button Step 3 Put REF model Eye on the head rest Firmly fix the KEF Setup tool with screws on the chin rest ...

Страница 3: ...Choose MODE 1 Select MODE content and press MODE button 2nd button Appearance of Figure 2 2 Moving machine head to find the most clear position The position which the rings looks most clearly is focusing position 3 Fixed machine head Figure 1 Figure 2 4 Adjust the position of the C CD camera until the picture is s hown in Figure 3 Appear 6 point Press 6th button adjust LED 50 Figure 3 ...

Страница 4: ...4 Do not need to tighten so that the CCD camera can move When you move the CCD camera pay attention to the C and Y value on the screen Other values can be ignored Adjust the camera process must be kept in the most clear position After the completion of save Figure 4 C 0 or C 1 Y 240 1 Note C and Y values must be specified otherwise it will cause CYL measure errors CCD camera position adjustment is...

Отзывы: