Sonicaid FM800 Encore Series Instructions for Use
68
8.3
75B
Responding to alarms
Type of alarm
Recommended user response
Signal outside range
Acknowledge alarm
Responsible clinician to decide what action to take
Loss of signal
Acknowledge alarm, if appropriate
For FHR: reposition the transducer
For other parameters: check transducer attachment and
connections
Timer
Depends on why the timer was set in the first place
FM800E failure
1 Stop monitoring.
2 Remove transducers from mother.
3 If there is an error message, make an exact note
of what it says.
4 Return FM800E to Huntleigh Healthcare Ltd or their
local representative.
212B
Acknowledging alarms
To acknowledge an alarm: > ALARM ACKNOWLEDGE in the main screen. This turns off
the audible alarm and resets the counters on which the calculation of alarm states is based.
The visual indication continues – i.e. the value which is out of range continues to flash. An
acknowledgement indicator will be printed identifying the start time of the alarm and the
time of the acknowledgement.
If more than one alarm has been triggered, ALARM ACKNOWLEDGE acknowledges them
all.
Содержание Sonicaid FM800 Series