Sonicaid FM800 Encore Series Instructions for Use
50
7
6B
Monitoring Maternal Parameters
7.1
67B
Contractions (using Toco transducer)
Use only Toco transducers supplied with the FM800E, or
listed in the FM800E Accessories & Consumables brochure.
1 Ensure that the Toco transducer and belt are clean and ready for use. In particular,
check the transducer for cracks or signs of damage. See also cleaning instructions in
section 17.
2 Connect the Toco transducer to the pink socket on FM800E.
3 Place the belt round the abdomen, and secure it over the transducer button so that it is
retained on the midline over the fundus of the uterus.
4 DO NOT USE COUPLING GEL. Wipe off any gel present on abdomen around this
area.
5 Contractions activity is measured as a % of full scale deflection. The contractions
measurement automatically zeroes to 10%. If it is suspected that the
transducer is not zeroed correctly, press the pink Toco Zero button on
the front of FM800E, at a moment when the mother is not
experiencing a contraction. Check the baseline periodically and re-zero the Toco if
necessary.