4
PREVENTIVE MAINTENANCE
Preventive maintenance should be carried out every 12 months.
4.1
Servicing Protocol
4.1.1
Procedure to Remove Top Cover
1. Ensure pump is switched Off and the power cord is removed from the
supply outlet.
2. Turn pump over and remove the four mounting feet using a flat blade
screwdriver.
3. Remove the four screws from the base using a suitable screwdriver.
4. Turn the pump over as one whole unit and then carefully remove the
top cover.
247901
4-1
Issue 1
March 1998
WARNING
Before dismantling the pump unit, ensure unit has been isolated
from the power supply by removing the cord plug from the wall
outlet.
CAUTION : Static Sensitive Devices
This pump should only be opened by personnel trained in ESD
methods and with appropriate equipment and anti-static
protection.
WARNING :
A potential ELECTRICAL SHOCK HAZARD exists when the top
cover is removed, EVEN with the pump switched off.
Содержание Flowtron AC550
Страница 1: ...247901 Issue 1 Mar 1998 AC550 Flowtron Excel system Service Manual Huntleigh HEALTHCARE...
Страница 2: ...247901 Issue 1 Mar 1998...
Страница 4: ...247901 ii Issue 1 March 1998...
Страница 6: ...247901 iv Issue 1 March 1998...
Страница 12: ...1 4 Pneumatic Schematic 247901 1 4 Issue 1 March 1998...
Страница 13: ...1 5 Valve Sequence 247901 1 5 Issue 1 March 1998...
Страница 14: ...1 6 Electrical Wiring Schematic 247901 1 6 Issue 1 March 1998...
Страница 15: ...1 7 General Assembly 247901 1 7 Issue 1 March 1998...
Страница 16: ...247901 1 8 Issue 1 March 1998...
Страница 18: ...247901 2 2 Issue 1 March 1998...
Страница 30: ...247901 4 8 Issue 1 March 1998...
Страница 32: ...5 4 Replacing Main PCB or Front Display Panel 247901 5 2 Issue 1 March 1998...
Страница 33: ...5 5 Replacing the Motor Gearbox Assembly 247071 247901 5 3 Issue 1 March 1998...
Страница 35: ...5 10 Pressure Control Assembly 247901 5 5 Issue 1 March 1998...
Страница 38: ...247901 5 8 Issue 1 March 1998...
Страница 40: ...247901 6 2 Issue 1 March 1998...