General
771402-2 dopplex ABIlity Service Manual
9
Recommended separation distances between portable and mobile RF communications
equipment and the dopplex Ability
The dopplex Ability is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled.
the customer or user of the dopplex Ability can help prevent electromagnetic interference by maintaining a minimum
distance between portable and mobile RF communications equipment (transmitters) and the dopplex Ability as
recommended below, according to the maximum output power of the communications equipment.
Rated maximum
output power of
transmitter
W
Separation distance according to frequency of transmitter
m
150kHz to 80MHz
d
= 1.2
√
P
80MHz to 800MHz
d
= 1.2
√
P
800MHz to 2.5GHz
d
= 2.3
√
P
0.01
0.12
0.12
0.23
0.1
0.38
0.38
0.73
1
1.2
1.2
2.3
10
3.8
3.8
7.3
100
12
12
23
For transmitters rated at a maximum output power not listed above, the recommended separation distance
d
in metres (m)
can be estimated using the equation applicable to the frequency of the transmitter, where
P
is the maximum output power
rating of the transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1 At 80MHz and 800MHz, the separation distance for the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection from structures, objects and people.
2.9
Electrostatic Discharge
Electrostatic discharge (ESD) is a known problem that can affect electrical equipment. If the dopplex Ability is
subjected to ESD during an ABI measurement, it is possible that the measurement will be suspended. The air in
all cuff chambers will be rapidly exhausted, and the unit will reset to the start-up screen.
If this happens, simply perform the usual checks and repeat the test.
2.10 Service & Warranty
Huntleigh Healthcare's standard terms and conditions apply to all sales. These contain full details of warranty
terms and do not limit the statutory rights of the consumer. A copy is available on request.
If there is any reason to return the dopplex Ability unit, please:
1.
Clean the product, as described in the cleaning section.
2.
Pack it in suitable packing.
3.
Attach the decontamination certificate (or other written statement declaring that the product has
been cleaned) to the outside of the package.
4.
Mark the package "Service Department – dopplex Ability” and return to the address below.
Huntleigh Healthcare Ltd reserve the right to return product that does not contain a decontamination
certificate.
Содержание DA100PB
Страница 1: ...771402 2 dopplex ABIlity Service Manual 1 SERVICE MANUAL ...
Страница 15: ...Product Description 771402 2 dopplex ABIlity Service Manual 15 ...
Страница 22: ...System Block Diagram 771402 2 dopplex ABIlity Service Manual 22 6 7 Pneumatic Circuit Diagram ...
Страница 23: ...System Block Diagram 771402 2 dopplex ABIlity Service Manual 23 ...
Страница 24: ...System Block Diagram 771402 2 dopplex ABIlity Service Manual 24 ...
Страница 29: ...Product Specifications 771402 2 dopplex ABIlity Service Manual 29 8 4 General ...
Страница 66: ...Fault Modes 771402 2 dopplex ABIlity Service Manual 66 Figure 3 Inflation Envelope ...