SpectraTrend HT User’s Guide 3.2
33
Choose
HITCH
when you wish to alter the readings made on the SpectraTrend HT to better
correlate to another color measurement instrument. A specific standard with known color
values from the reference instrument are read with the SpectraTrend HT. The reading is stored
in the product setup with the known color values from the reference instrument. In this way
the instruments are hitched together. The modified standard is stored in the product setup for
as long as desired.
Figure 18. Product Setup: Hitch
Choose
PHYSICAL
when an actual product specimen is available and represents your target
color. Use the target color values of the product specimen to compare to the color values of
your samples. A physical standard is measured and stored in the product setup for as long as
desired.
Figure 19. Product Setup & Physical Standard Type
Average
The average is only available for
READ
Mode. When using the
RUN
mode, this option setting is
ignored.
Select
OFF
if you wish to make only one reading of each standard and sample when using the
manual
READ
mode. If you would instead like to average several readings into a single
measurement when using the
READ
mode, indicate the number of readings to be averaged, up
to 20.
Press the
RIGHT ARROW
button to move through the list of available choices.
Содержание SpectraTrend HT
Страница 6: ...SpectraTrend HT User s Guide 3 2 6...
Страница 21: ...SpectraTrend HT User s Guide 3 2 21...
Страница 22: ...SpectraTrend HT User s Guide 3 2 22...
Страница 23: ...SpectraTrend HT User s Guide 3 2 23...
Страница 24: ...SpectraTrend HT User s Guide 3 2 24...
Страница 25: ...SpectraTrend HT User s Guide 3 2 25...
Страница 26: ...SpectraTrend HT User s Guide 3 2 26...
Страница 27: ...SpectraTrend HT User s Guide 3 2 27...
Страница 28: ...SpectraTrend HT User s Guide 3 2 28...
Страница 76: ...SpectraTrend HT User s Guide 3 2 76...
Страница 86: ...SpectraTrend HT User s Guide 3 2 86...
Страница 92: ...SpectraTrend HT User s Guide 3 2 92...