ColorFlex EZ Citrus Citrus
User’s Manual Version 2.0
15
14.
Press
UP
to move to the next data view for this product setup.
15.
Press
RIGHT ARROW
to retur to the main menu.
Using a Hitch to the Citrus Standard
The purpose of the Citrus Standard is to improve inter-instrument agreement in Citrus Score
values at multiple sites, or between a buyer and seller of processed Citrus products. The hitch
standard is particularly important for the ColorFlex Citrus because
in the first setup (“
Citrus
Scores”) hitches the instrument to the assigned values of the
Citrus Standard.
At the factory, the values for the OJ4 Standard (C/2°) have been entered along with the product
views. If it is necessary to re-set the hitch, here is the procedure.
1.
From the
MAIN MENU
, choose
PRODUCT SETUP
.
Figure 15. Yes to Select Setup
2.
Select the
CITRUS SCORES SETUP
by highlighting the selection and pressing
GO
.
Figure 16. Select Citrus Scores
3.
Make sure that the
STANDARD TYPE
is set to
HITCH
; the
AVERAGE
is set to
OFF
, then select
the
VIEWS
.
Figure 17. Product Setup
4.
For each Citrus Score, the
VIEW
is set to Absolute, the Observer to C/2°, and the scale to
X,Y,Z
Содержание ColorFlex EZ Citrus
Страница 6: ...ColorFlex EZ Citrus Citrus User s Manual Version 2 0 6...
Страница 18: ...ColorFlex EZ Citrus Citrus User s Manual Version 2 0 18...
Страница 44: ...ColorFlex EZ Citrus Citrus User s Manual Version 2 0 44...
Страница 54: ...ColorFlex EZ Citrus Citrus User s Manual Version 2 0 54 Regulatory Notice...
Страница 60: ...ColorFlex EZ Citrus Citrus User s Manual Version 2 0 60...