13
7.
Press the button
GO
. The ColorFlex EZ Citrus reads the White Tile and sets the top-
of-scale. When it is finished, the screen indicates that the instrument has been
successfully standardized.
8.
Press
MAIN MENU
(right arrow) to return to the main menu.
9.
Replace the standard port plate with the Citrus tube holder.
10.
Periodically (i.e. once per week), perform the
White Tile
check and
Green Tile
check
to ensure that the instrument is operating properly. The calibrated Citrus red tile
should also be read to confirm that the instrument hitch is valid.
Reading Individual Samples
Complete the following steps to take individual readings using the ColorFlex EZ Citrus:
Note: These instructions apply when Average in the product setup is set to OFF.
1.
From the
MAIN MENU
, press the
DOWN ARROW
key until the product setup screen
is reached. Press
GO
to select a setup.
Figure 11. Main Menu > Product Setup
2.
Press the
UP
or
DOWN ARROW
to select the
CITRUS COLOR SETUP
or
OTHER
JUICES
.
Press
GO
to continue.
Figure 12. Select Setup